Accelerated Aging
Calculate the duration of storage tests
Work out how long a storage test at elevated temperature needs to run in order to reflect a real storage period in the laboratory as closely as possible. The calculation is based on the Arrhenius relationship with Q10 = 2 (standard requirement).

Free tool
Enter your values
Enter the real storage temperature, the duration to be simulated and the planned test temperature – the acceleration factor and the test duration appear instantly.
Note:
Please note that this is only a rough guideline value and that we give no warranty for this free service. The result should always be checked for each individual case.
How the calculation works
According to the Arrhenius relationship, aging processes speed up as temperature rises. With Q10 = 2, the reaction rate is assumed to double for every 10 °C increase in temperature.
AAF = Q10(TAA − TRT) / 10
Acceleration factor from Q10 and the temperature difference
Test duration = duration to be simulated / AAF · TAA = test temperature, TRT = real storage temperature
Worked examples
Example 1
365 days at 23 °C
≙ 39.7 days at 55 °C (AAF 9.19)
Example 2
365 days at 23 °C
≙ 79.4 days at 45 °C (AAF 4.59)
Example 3
730 days at 25 °C
≙ 129 days at 50 °C (AAF 5.66)
Putting accelerated storage tests into perspective
We recommend verifying key results from accelerated storage tests with storage tests under real conditions. The elevated temperature can lead to aging effects being overestimated; because of the shortened test duration, underestimates are also possible. Nevertheless, accelerated storage tests are indispensable for estimating time-dependent aging effects and as a basis for risk management decisions.
Storage tests in the laboratory
Accelerated aging at Innoform
We carry out storage tests and accelerated aging in heating or cooling cabinets with controlled temperature, in climate chambers with controlled temperature and relative humidity, and in explosion-proof heating chambers – and then test your materials for the properties that matter to you.
If you have questions about the calculator or about planning your storage test, please get in touch.
Your contact person

Dr. Daniel Wachtendorf
+49 441 94986-22
daniel.wachtendorf@innoform.de
Top Tests in This Area
Test Method Search
Puncture resistance, quasi-statical (temperature range – 40 up to + 120 °C)
Mechanical propertiestensile test films and laminates (temperature range -40 to +120 °C)
Mechanical properties- Mechanical properties
- Mechanical properties
- Mechanical properties
Seal strength, tensile test on strips (temperature range -40 to +120 °C)
Mechanical properties- Mechanical properties
Sealing and welding properties of films, verification and comparison by laboratory sealing press
Mechanical propertiesDetermination of thickness of plastic films, aluminium foils, plane multilayer-materials and sheets
Mechanical properties- Mechanical properties